SECTION 7 PERFORMANCE TEST

7-9

(4) Test procedure

STEP

PROCEDURE

FM deviation

1.

Set the MG3641A/MG3642A output level to +7 dBm.

2.

Set the MG3641A/MG3642A frequency and modulation analyzer receive frequency to the
expected frequency.

3.

Turn on the MG3641A/MG3642A FM, and set Source to Int1 1 kHz.

4.

Set the MG3641A/MG3642A FM frequency deviation.

5.

Read the indicator of the modulation analyzer.

6.

Change the MG3641A/MG3642A FM frequency deviation and repeat the measurement.

FM distortion

7.

Set the MG3641A/MG3642A FM frequency deviation to 3.5 kHz or 22.5 kHz.

8.

Measure the demodulation output of the modulation analyzer using the distortion meter.

(5) Note on test

• The demodulation zone of the modulation analyzer must be within 0.3 to 3 kHz at measurement of the FM

deviation and 0.3 to 15 kHz or 0.3 to 20 kHz at measurement of distortion rate.

• Using a modulation analyzer with much residual FM affects the measurement of distortion if the FM deviation

rate is lower. Use a modulation analyzer with a little residual FM.