2 measuring the 1-signal selectivity – Anritsu MG3641A User Manual

Page 96

Advertising
background image

SECTION 5 MEASUREMENT

5.2

Measuring the 1-signal Selectivity

The 1-signal selectivity is measured when the receiver operates in a proportional region because the expected wave

and interference disturbing wave are fine. It is indicated by the relative input voltage rate required to equalize the

output of the receiver by connecting the signal generator to the receiver input pin in the expected wave receiving state

and changing its frequency to the expected or disturbing one. This method is available to measure a pass band width,

attenuation slope, and spurious response.

5.2.1

Measuring selectivity characteristics of the FM receiver in 20 dB NQ method

This figure shows the selectivity characteristics of a

signal-channel FM receiver 146 to 162 MHz. The

approved values are as follows:

• Pass band width ........ 6 dB (Reduction width = 20

kHz or more)

• Attenuation slope ...... 70 dB (Reduction band width =

50 kHz or less)

These conditions apply to measure the selectivity in the

20 dB NQ method as follows:

• Pass band width ........ Increase the SG output by 6 dB

as compared with the NQ

sensitivity and move up or

down the frequency so that the

NQ sensitivity is obtained

again. Then, obtain this width

from its frequency width.

• Attenuation factor ..... Increase the SG output by 60

dB, not 6 dB like the above.

• The characteristics curve must not be in the hatched part. The solid line is approved; the broken line is rejected.

;

;

;

;

;

;

;

;

80

70

60

50

40

30

20

10

6

;

;

;;

;;

10

20

30

30

10

20

+kHz

–kHz