SECTION 6 GPIB

6-57

6.10.2 Restrictions in MG3633A command interchange mode

This paragraph describes the restrictions in MG3633A command interchange mode.

(1) Common items

• Exponent representation of numeric data is unavailable.

• The semicolon (;) is the only valid separator when multiple commands are sent on one line. Do not use the

comma (,).

• No header is added to talker data for a query command. Only numeric data is output.

(2) Frequency

• Frequency up and down (TFR and EFR) using the rotary knob are enabled by substituting the step key func-

tion. Using the step key and rotary knob to increase or decrease the frequency may not perform correctly

unless a frequency step value (FIS<Numeric><Freq term>) or frequency resolution (R0 to R9) is set.

• Resetting the frequency (ZFR) cannot be interchanged because the MG3641A/MG3642A does not have an

equivalent function.

• The frequency offset display function can be interchanged only with the multistatement description

(FOS<Numeric><Freq term>;SP12) of the offset frequency setting command (FOS<Numeric><Freq term>)

and offset display mode ON command (SP12) separated by a semicolon (;).

(3) Output level

• Output level up and down (TOL and EOL) using the rotary knob are enabled by substituting the step key func-

tion. Using the step key and rotary knob to increase or decrease the level may not perform correctly unless the

level step value (OIS<Numeric><Ampl term>) or level resolution digits (L0 to L2) are set.

• The level resetting function (ZOL) cannot be interchanged because the MG3641A/MG3642A does not have

an equivalent function.

• The level offset display function can be interchanged only with the multistatement description

(OOS<Numeric><Ampl term.;SP08) of the offset level setting command (OOS<Numeric><Ampl term>) and

offset display mode ON command (SP08) separated by a semicolon (;).

• If the output level setting (OL<Numeric><Ampl term>) or level step up or down (UOL, DOL) command is

executed in level continuous mode, the continuous mode is canceled.

• The command relating to the output level limiter (OLM<Numeric><Ampl term>, SP05, SP06) cannot be

interchanged because the MG3641A/3642A does not have an equivalent function.