3 connecting the gpib cable, Caution – Anritsu MG3641A User Manual

Page 120

Advertising
background image

SECTION 6 GPIB

6-14

6.3

Connecting the GPIB Cable

A GPIB cable must be connected for remote control by the GPIB. The MG3641A/MG3642A supports the GPIB

cable connector on the rear panel.

CAUTION

Turn the POWER switch off and unplug the power cord before connecting and disconnecting the GPIB cable.

This is because only the signal common line of the cord may rarely be cut before the others depending the cord

connection-disconnection method. In this case, if the power remains turned on, the AC leak voltage, etc. is

superimposed to the IC, and the circuit parts may be damaged.

A system using GPIB is restricted as follows. The device must be connected according to the following conditions:

• Number of connectable devices

≤15

• Total cable length

≤2 m × Number of devices (max. 20 m)