SECTION 7 PERFORMANCE TEST

7-2

Power meter

7.2

Performance Test Device List

Table 7-1 lists the measuring instruments for calibration according to the test parameters.

Table 7-1. Performance Test Device List

Test parameter

Measuring instrument

Requested

Recommended

performance*

device

Output frequency

Frequency counter

100 kHz to 2.08 GHz

MF1603A

Output level frequency

100 kHz to 2.08 GHz

ML4803A

characteristics

–30 to +20 dBm

MA4601A

Spectrum Analyzer

100 kHz to 2.08 GHz

MS2602A

Output level accuracy

Pre-amplifier

100 kHz to 2.08 GHz
40 dB

FM frequency deviation

Modulation analyzer

100 kHz to 2.08 GHz

MS616B

AM modulation factor

AM, FM

Modulation distortion

Distortion meter

20 Hz to 100 kHz

* Extracts a part of performance that can cover the test parameter measurement range.