SECTION 7 PERFORMANCE TEST

7-5

7.3.2

Output level frequency characteristics

(1) Test specifications

• Frequency characteristics

with reference to 0 dBm
±0.5 dB

(2) Measuring instrument for test

• Power meter (100 kHz to 2080 MHz)

(3) Setup

Fig. 7-2. Output Level Frequency Characteristic Test

(4) Test procedure

STEP

PROCEDURE

1.

Adjust the power meter to the zero point and correct the sensor sensitivity.

2.

Set the MG3641A/MG3642A output level to 0 dBm.

3.

Set the MG3641A/MG3642A to the frequency to be measured.

4.

Set the calibration factor of the sensor of the power meter and read the output level.

5.

Repeat steps 3 and 4.

(5) Note on test

The MG3641A/MG3642A output level is defined with the RF output connector. To measure the output level,

directly connect the power sensor to the MG3641A/MG3642A RF output connector.

Sensor

Power meter

SENSOR
INPUT

MG3641A/MG3642A