4 level continuous mode – Anritsu MG3641A User Manual

Page 54

Advertising
background image

SECTION 4 OPERATING INSTRUCTIONS

4.5.4

Level Continuous Mode

This signal generator employs a mechanical attenuator to vary the output level. For this reason, momentary signal

interruptions and spike noise may be produced when you vary the output level.

Therefore, you select the level continuous mode in the cases where such momentary signal interruptions and spike

noise pose particular problems in the measurements you are conducting.

When you activate the level continuous mode, the operation of the mechanical attenuator will be fixed, and you will

be able to change the level continuously within a maximum range of

±10 dB by using only the electronic attenuator

designed for high-resolution setting.

(1) Setting the Level Continuous Mode

Press the [Shift] key and, then, the [

∨] (Continuous) key in the Output Level zone. The characters "Cont" will

appear on the Output Level indicator, indicating that the continuous mode has been set up.

The output level setting range will be limited to

±10 dB with its center at the level that existed when the continuous

mode was activated, but you can vary the level within that range, free from momentary signal interruptions or

spike noise.

(2) Clearing the Level Continuous Mode

Press the [Shift] key and, then, the [^] (Normal) key in the Output Level zone. The characters "Cont" will go out

on the Output Level indicator, indicating that the continuous mode has been cleared.

MG3641A/MG3642A front panel

[

] (Normal) key

[

] (Continuous) key

[Shift] key