SECTION 5 MEASUREMENT

5.2.2

Measuring spurious response

The spurious sensitivity reduces as the receiver output obtained at reception of the modulated desired-frequency is

different from that obtained at reception of the modulated spurious frequency. To measure the spurious response,

adjust the output level of the signal generator of the spurious frequency so that these two receiver outputs are equal.

Then, obtain that value from the difference between the output level of the signal generator of the desired frequency

and that of the signal generator of the spurious frequency.

When the desired frequency of the receiver is fd; IF frequency is fi; local frequency is fL, the following frequency

is assumed to be spurious frequency fs.

• Image frequency ......... fs = fL

±fi = fd ±2fi

• Higher harmonic ......... fs = fL

±fi/2, fs = nfd ±fi/2

If a frequency that sets the difference from the local frequency to fi/2 is input, the second

higher harmonic of fi/2 changes to an IF frequency and an interference occurs.

• Harmonic frequency of local frequency ....... fs = nfL

±fi

This section shows an example of fd = 154.45 MHz, fs = fd + 2 fi. fi = 10.7 MHz to explain the measurement

procedure.

(1) Setup

Fig. 5-4. Measuring Spurious Sensitivity

MG3641A/MG3642A

DA

FM RX

DL

LM

(154.45MHz)

Pseudo-

antenna

FM

receiver

Standard

pseudo-load

Level

meter