Section 7. Measurement Instructions

7-9

BrFull6W (Dest, Reps, Range1, Range2, DiffChan, ExChan, MeasPEx,

ExmV, RevEx, RevDiff, SettlingTime, Integ, Mult, Offset)

X = result w/mult = 1, offset = 0

X

V

V

R

R

R

R

R

R

=

=

+

+



1000

1000

2

1

3

3

4

2

1

2


This Instruction applies an excitation voltage and makes two differential
voltage measurements. The measurements are made on sequential channels.
The result is the voltage measured on the second channel (V2) divided by the

voltage measured on the first (V1). The result is 1000 times V2 / V1 or

millivolts output per volt of excitation. The connections are made so that V1 is

the measurement of the voltage drop across the full bridge, and V2 is the

measurement of the bridge output.

Full Bridge Parameters

Parameter
& Data Type

Enter

Dest

Variable or
Array

The Variable in which to store the results of the instruction. When Reps are used the
results are stored in an array with the variable name. An array must be dimensioned to
have elements for all the Reps.

Reps

Constant

The number of repetitions for the measurement or instruction.

Range

Constant

Alpha Code

Numeric
Code

Voltage
Range

mV5000 0

± 5000 mV

mV1000 1

± 1000 mV

mV200 2

± 200 mV

mV50 3

± 50 mV

mV20 4

± 20 mV

AutoRange 5

mV20 - mV5000 Selects range (Sect. 3.1)

mV200C

mV50C
mV20C

20
30
40

± 200 mV

± 50 mV
± 20 mV

The mV200C, mV50C, and mV20C
ranges pull the channel into common
mode range and check for open input

DiffChan

Constant

The differential channel number on which to make the first measurement. When Reps are
used, subsequent measurements will be automatically made on the following differential
channels. If the channel is entered as a negative number, all reps occur on that channel.