Section 6. Data Table Declarations and Output Processing Instructions

6-5

DataEvent Example:

In this example, 5 type T thermocouples are measured. The trigger for the start
of an event is when TCTemp(1) exceeds 30 degrees C. The stop trigger is
when TCTemp(1) less than 29 degrees C. The event consists of 20 records
prior to the start trigger and continues to store data until 10 records following
the stop trigger.

Const

RevDiff 1

'Reverse input to cancel offsets

Const

Del 0

'Use default delay

Const

Integ 0

'Use no integration

Public

RefTemp

Declare the variable used for reference temperature

Public

TC(5)

'Declare the variable used for thermocouple measurements

Public

Flag(8)

Units

RefTemp=degC

'

Units

TC=degC

DataTable

(Event,1,1000)

DataInterval

(0,00,msec,10)

'Set the sample interval equal to the scan

DataEvent

(20,TC(1)>30,TC(1)<29,10)

'20 records before TC(1)>30,

'after TC(1)<29 store 10 more records

Sample

(1,RefTemp,IEEE4)

'Sample the reference temperature

Sample

(5,TC,IEEE4)

'Sample the 5 thermocouple temperatures

EndTable

BeginProg

Scan

(500,mSec,0,0)

PanelTemp

(RefTemp,250)

TCDiff

(TC(),5,mV50C,1,TypeT,RefTemp,RevDiff,Del,Integ,1,0)

CallTable

Event

NextScan

EndProg

FillStop

Data Tables are by default ring memory where, once full, the newest data are
written over the oldest. Entering FillStop into a data table declaration makes
the table fill and stop. Once the table is filled, no more data are stored until the
table has been reset. The table can be reset from within the program by
executing the ResetTable instruction. Tables can also be reset from PC9000
real time windows or the collect data window.

Example:

DataTable (Temp,1,2000)

DataInterval(0,10,msec,10)
FillStop

' the table will stop collecting data after 2000 records.

Average(1,RefTemp,fp2,0)
Average(6,TC(1),fp2,0)

EndTable