3 effect of grounding on single-ended measurements, 8 powering sensors and peripherals – Campbell Scientific CR5000 Measurement and Control Module User Manual

Page 33

Advertising
background image

Section 1. Installation and Maintenance

1-9

1.7.3 Effect of Grounding on Single-Ended Measurements

Low-level single-ended voltage measurements can be problematic because of
ground potential fluctuations. The grounding scheme in the CR5000 has been
designed to eliminate ground potential fluctuations due to changing return
currents from 12 V, SW-12, 5 V, and the control ports. This is accomplished
by utilizing separate signal grounds (

) and power grounds (G). To take

advantage of this design, observe the following grounding rule:

Always connect a device’s ground next to the active terminal
associated with that ground.

Examples:

1.

Connect 5 Volt, 12 Volt, and control grounds to G terminals.

2.

Connect excitation grounds to the closest

terminal on the excitation

terminal block.

3.

Connect the low side of single-ended sensors to the nearest

terminal on

the analog input terminal blocks.

4.

Connect shield wires to the nearest

terminal on the analog input

terminal blocks.

If offset problems occur because of shield or ground leads with large current

flow, tying the problem leads into the

terminals next to the excitation, CAO,

and pulse-counter channels should help. Problem leads can also be tied
directly to the ground lug to minimize induced single-ended offset voltages.

1.8 Powering Sensors and Peripherals

The CR5000 is a convenient source of power for sensors and peripherals
requiring a continuous or semi-continuous 5 VDC or 12 VDC source. The
CR5000 has 2 continuous 12 Volt (12V) supply terminals, 2 switched 12 Volt
(SW-12) supply terminals, and 2 continuous 5 Volt (5V) supply terminals.
Voltage on the 12V and SW-12 terminals will change with the CR5000 supply
voltage. The 5V terminal is regulated and will always remain near 5 Volts
(

±

4%)so long as the CR5000 supply voltage remains above 11 Volts. The 5V

terminal is not suitable for resistive bridge sensor excitation. Table 1.8-1
shows the current limits of the 12 Volt and 5 Volt ports. Table 1.8-2 shows
current requirements for several CSI peripherals. Other devices normally have
current requirements listed in their specifications. Current drain of all
peripherals and sensors combined should not exceed current sourcing limits of
the CR5000.

NOTE

Advertising