12) menu p: parameter protection items, 12) menu p: parameter protection items -34 – APC AXFA11G User Manual

Page 62

Advertising
background image

IM 01E20C01-01E

6-34

6. PARAMETER DESCRIPTION

(3) “Normal” will be restored when the power is

turned off or when 30 minutes have elapsed
since entry to Test Mode.

(4) In Test Mode, the warning 83: Fix Cur Wng

will be displayed as a warning message. (For
more details, refer to Section 6.5 Alarm
Functions.)

[N11: Test Output Value]

Setting for test output

values
During loop testing, current output (4 to 20mA),
totalization, and pulse will be output in accordance
with this parameter’s setting, and values can be set
when “Test” has been selected for

N10: Test Mode

.

With multiple ranges or when performing forward/
reverse flow measurements, setting should be done
using a percentage of the maximum range.

[N20: Test SO1]

Setting for SO1 output terminal

condition during testing
This parameter sets the condition of the SO1 status
output terminal during loop testing. Setting is possible
when “Test” has been selected for

N10: Test Mode

.

Setting

Function

SO1 output terminal in Open (Off) condition

SO1 output terminal in Closed (On) condition

Open (Off)

Closed (On)

T0645.EPS

[N21: Test SO2]

Setting for SO2 output terminal

condition during testing
This parameter sets the condition of the SO2 status
output terminal during loop testing. Setting is possible
when “Test” has been selected for

N10: Test Mode

.

Setting

Function

SO2 output terminal in Open (Off) condition

SO2 output terminal in Closed (On) condition

Open (Off)

Closed (On)

T0645-2.EPS

[N22: Test Alarm Out]

Setting for alarm terminal

condition during testing
This parameter sets the condition of the alarm output
terminal (AL) during loop testing.

Setting

Function

AL output terminal in Open (Off) condition

AL output terminal in Closed (On) condition

Open (Off)

Closed (On)

T0645-3.EPS

[N23: Test SI1]

Display for SI1 status input terminal

condition during testing
This parameter is used to display the condition of the
SI1 status input terminal during loop testing.

Setting

Function

SI1 input terminal in Open condition

SI1 input terminal in Short condition

Open

Short

T0646.EPS

[N24: Test SI2]

Display for SI2 status input terminal

condition during testing
This parameter is used to display the condition of the
SI2 status input terminal during testing.

Setting

Function

SI2 input terminal in Open condition

SI2 input terminal in Short condition

Open

Short

T0646-2.EPS

(12) Menu P: Parameter Protection items

Menu P contains items that are relevant to write
protection and passwords.

[P10: Key Code]

Parameter of the display restriction

This parameter restricts access to the Service Mode.

NOTE

Write Protect function

᭿

The parameters P20 through P23 are set
when using the write protect function. Specifi-
cally, this function responds to a hardware
switch or the setting of a software password,
and it protects parameters from being over-
written.

᭿

If the hardware switch is set to “Protect”, it will
not be possible to overwrite parameters;
furthermore, this condition will be maintained
until the switch is set to “Enable”.

᭿

For more details regarding hardware switch
settings, refer to Section 10.2.2: Setting of
Write Protect Switch.

[P20: Write Protect]

Status indication for write

protection
This parameter is used to indicate whether or not write
protection is currently on.

F0620.EPS

Default setting (Enable)

P20:Write Protect

No

P20:Write Protect

Yes

F0621.EPS

Write protection (Protect)

Advertising