Hot key 2. wrt protect menu 2. enable wrt 10min, Open “ wrt protect menu ” in the hot key menu, 6 7 8 9 a – APC AXFA11G User Manual

Page 93

Advertising
background image

IM 01E20C01-01E

8-19

8. OPERATION VIA HART COMMUNICATOR

Changing Password

Example: Change the password from “1 2 3 4” to “6 7 8 9 A”.

F0822-1.EPS

ABORT

ABORT

OK

AXFA11 : FI-1201
Enter current
password to enable to
write for 10
minutes:

Hot key

2. Wrt Protect Menu

2. Enable Wrt 10min

HELP

DEL

ABORT

ENTER

1

2

AXFA11 : FI-1201
If you wish to
release completely,
you have to change
password to all of
spaces.

HELP

DEL

ABORT

OK

3

Open “Wrt protect menu” in the Hot key Menu.

AXFA11 : FI-1201
Release the write
protection for 10
minutes.

DEL

DEL

ABORT

OK

AXFA11 : FI-1201
Wrt Protect Menu
1 Write protect No
2 Enable Wrt 10min
3 New Password
4 Software seal Keep

HELP

SAVE

ABORT

OK

4

5

AXFA11 : FI-1201
Enter new password to
change state of write
protect

HELP

DEL

ABORT

ENTER

F4

(OK)

F4

(OK)

F4

(ENTER)

YZ/

3

6

AXFA11 : FI-1201
Re-enter new Password
write 30 seconds:
6789A
6789A

HELP

DEL

ABORT

ENTER

7

AXFA11 : FI-1201
Change to New
password

HELP

DEL

F4

8

(OK)

AXFA11 : FI-1201
It changed the state
of protection related
password.

HELP

DEL

OK

F4

(ENTER)

“1 2 3 4”

Enter the password and press ENTER (F4).

Press OK (F4).
Write protect status is released for 10 minutes.

Press OK (F4).

Select “New Password”.

Enter “6 7 8 9 A” and press ENTER (F4).

Enter “6 7 8 9 A” again and press ENTER (F4).

Press OK (F4).

“6 7 8 9 A”

F4

(ENTER)

“6 7 8 9 A”

F0822-2.EPS

or

Advertising