8. OPERATION VIA HART COMMUNICATOR

NOTE

Data changed with the HART Communicator is sent to the AXFA11 by pressing

SEND (F2)

of the HART Communicator.

8.3.3

Checking for Problems

The self-diagnostic function of the AXFA11 is explained in Section 6.5 “Alarm Functions”
By using the HART Communicator, it is also possible to carry out this function in the
“Test/Status” parameter. Test for each error.

1. Device Setup

2. Diag/Service

3. Test/Status

* Open the “Test/Status” setting display.

8.3.4

Setting Parameters

As mentioned in Section 5.2.2 “Setting Mode,” the AXFA11’s parameters are divided into
three types:

i.

Selection type

ii.

Numeric type

iii.

Alphanumeric type

This section describes how to set these parameters using a HART communicator.

NOTE

All three parameters must be set to obtain a correct signal.
Nominal size, flow rate span and meter factor must be set.

8.3.4.1

Example of Setting Selection-Type Data: Base Flow Unit

This example describes the setting of the flow rate units for the selection-type parameter
“Base Flow Unit” from m

3

to l (Liter).

1. Device Setup

3. Easy Setup

3. Base Flow Unit

AXFA11 : FI-1100
Base Flow Unit
m3
m3
K1(Kiloliter)
l(Liter)
cm3

DEL

DEL

ESC

ENTER

1

3

Select “1(Liter)”.

Press ENTER (F4).

Press SEND (F2) to set the data into the AXFA11.

Open the “Base Flow Unit” setting display.

AXFA11 : FI-1100
Easy Setup
1 Language
2 PV Damping
3 Base Flow Unit l
4 Base Time Unit /h
5 PV Span

DEL

DEL

HOME

2

F4

(ENTER)

F2

(SEND)

×

2

F0807.EPS

SEND