1-3

1. INTRODUCTION

1.3 Combination Remote

Flowtubes

IMPORTANT

• The AXFA11 Magnetic Flowmeter Converter

should be used in combination with the follow-
ing remote flowtubes:
AXFA11G

AXF002

ٗ

-N to AXF26L

ٗ

-N

Contact Yokogawa before using it in combina-
tion with flowtubes other than those listed
above.

• The model AXF

ٗٗٗ

C remote flowtube with

optional code JF3 (TIIS flame proof type)
cannot be combined with the AXFA11 con-
verter. In this case, use the AXFA14 converter.

• If the converter combined with the AXF mag-

netic flowmeter remote flowtube is changed
from the AXFA11 to AXFA14 or vice versa, the
meter factor of the remote flowtube must be
readjusted according to its flow calibration.

CAUTION

In case of combination with the explosion proof
type remote flowtube (AXF

ٗٗٗ

C-N) for

CENELEC ATEX, FM, and CSA certification,
please see the manual IM 01E20D01-01E. The
construction of the instrument, installation,
external wiring, maintenance, and repair are
strictly restricted, and non-observance or negli-
gence of these restriction would result danger-
ous condition.