Using the Agilent E1439

Fiber Optic Interface

Fiber Modes

The E1439D’s fiber interface can operate in five different modes:

“Off” on page 45

“Copy” on page 46

“Raw” on page 47

“Generate” on page 48

“Append” on page 50

Off

The Off fiber mode disables the fiber transmitter but allows the fiber receiver to read control
signals. Normal data collection and filtering continues, and the data port selection determines
whether data is sent to the local bus (Agilent E1439D only) or read from the FIFO via the VME
bus. See the following illustration.

Note

Setting the data port to Fiber while in the Off fiber mode causes the data FIFO to fill up with
filtered ADC data, which then causes data collection to stop.

Fiber Interface Setup

Fiber Mode

Off

Rate

setting ignored

BOF

setting ignored

CRC

setting ignored

Flow Control

setting ignored

Epoch Generate

setting ignored

Epoch Size

setting ignored

Fiber TX

Fiber RX

VME BUS

LBUS

ADC

DIGITAL

FILTERS

FIFO

E1438D / E1439D