Digitale Eingänge (TTL-Pegel)

Digital inputs (TTL-level)

Entrées numériques (niveau TTL)

Pin 2, remote

Pin 3, start

Pin 6, input 1

Pin 8, input 2

Pin 13, speed intern

Analoge Eingänge

Analog inputs

Entrées analogiques

Pin 5, speed IN

0 – 5 V

DC

/ 0 – 10 V

DC

0 – 20 mA / 4 – 20 mA

Universal Ausgänge (PWM)

Universal outputs (PWM)

Sorties universelles (PWM)

Pin 14, output 1
Pin 15, output 2

Analoge Ausgänge

Analog outputs

Sorties analogiques

Pin 9, speed OUT

0 –10 V

DC

/ 0 – 12 kHz

8 7 6 5 4 3 2 1

15 14 13 12 11 10 9

speed IN

direction

input 1

start

+26VDC

remote

input 2

GND

output 2

speed OUT

output 1

+5VDC

speed intern

+39VDC

Analogschnittstelle

Pin 1, GND (Masse)
Bezugspotential für alle anderen
Eingänge.

Pin 2, remote
Für Umschaltung zwischen ma-
nueller Bedienung und der Ana-
logschnittstelle. Zur Aktivierung
der Analog-Schnittstelle muss
Pin 2 mit Pin 1 (GND) verbunden
werden.

Pin 3, start
- Im Normalbetrieb (Pin 2 offen)

dient Pin 3 als Fußschalter
(siehe DIP-Switch Einstellungen
Seite 35)

- Im Remote-Betrieb (Pin 2 auf

GND) startet die Pumpe bei
Verbindung mit Pin 1 (GND)

Pin 5, speed IN
Für externe Drehzahlsteuerung
(0 – 5 V, 0 – 10 V, 0 – 20 mA,
4 – 20 mA) Eingangsimpedanz
und Wahlmöglichkeit mittels DIP-
Switch im Geräteinnern (siehe
Seite 35)

Pin 6, input 1
Pin 8, input 2
Digitale Eingänge (TTL-Pegel).
Aktiv, wenn mit Pin 1 verbunden.
Können nur über RS232-Schnitt-
stelle bedient werden; z.B. mit
einer Software wie ProgEdit .

Pin 7, +39 V

DC

Es stehen ca. +39 V

DC

zur Verfü-

gung (max. Strom 1A).

Analog interface

Pin 1, GND (ground)
Reference potential for all other
inputs

Pin 2, remote
For changing between manual
control and analog interface. For
activating the analog interface,
pin 2 must be connected with pin
1 (GND).

Pin 3, start
- In normal operation (pin 2

open) pin 3 operates the
foot-switch
(see DIP switch settings,

page 35)
- In remote operation (pin 2 to

GND) the pump starts when
connected to pin 1 (GND)

Pin 5, speed IN
External speed control
(0 – 5 V, 0 – 10 V, 0 – 20 mA,
4 – 20 mA) Input impedance and
input range can be selected via a
dip-switch inside the pump (see
page 35)

Pin 6, input 1
Pin 8, input 2
Digital inputs (TTL-level). Active
when connected to pin 1. Are
only retrievable via RS232 inter-
face; e.g. with a software like the
ProgEdit.

Pin 7, +39 V

DC

About +39 V

DC

are available (max.

current 1A).

Interface analogique

Pin 1, GND (masse)
Potentiel de référence pour toutes
les autres entrées

Pin 2, remote
Pour commuter du service manuel
à l’interface analogique.
Pour activer l'interface analo-
gique, le pin 2 doit être connecté
au pin 1 (GND).

Pin 3, start
- En exploitation normale (pin 2
ouvert), le pin 3 sert d’inter-
rupteur au pied (voir réglages
des DIP-switch en page 35)
- en exploitation à distance (pin
2 sur GND), la pompe se met
en route dès qu’elle est
connectée au pin 1 (GND)

Pin 5, speed IN
Réglage externe du nombre de
tours (0 – 5 V, 0 – 10 V, 0 – 20
mA, 4 – 20 mA) Impédance
d‘entrée et réglage de zone au
moyen de l’interrupteur DIP à
l’intérieur de l’appareil (v.p. 35)

Pin 6, input 1
Pin 8, input 2
Entrées numériques (niveau TTL).
Actifs si reliés au pin 1. Ne peu-
vent être interrogés que par
l’interface RS232, p.e. avec le
logiciel ProgEdit.

Pin 7, +39 V

CC

Environ +39 V

CC

sont à disposi-

tion (courant maximal 1A).