Evaluation

Eppendorf SOP

English (EN)

3

Evaluation

3.1

Calculating the systematic error

Mean value of the dispensed volume:

n = Number of measured values

To convert the measured values into volume values, use correction factor Z for
the dependency of the test liquid on temperature and air pressure for each
individual value (see Factor Z for distilled water on p. 36).

Systematic error e

s

in micro liter:

Systematic error e

s

in percent:

3.2

Calculating the random error

Random error as repeat standard s:

Random error as coefficient of variation CV:

The specifications of the tested pipette can be found in the relevant operating
instructions or technical specifications (see Technical specifications on p. 37).