Adjustment
Eppendorf SOP
English (EN)

7.2.5

Research variable

Auxiliary equipment
• Supplied pipette wrench (order no. 3111 501.016)
• Supplied red CAL seal

4. A revolution corresponds to:

5. Pull off the pipette wrench.
6. Move the volume setting ring slightly forwards and backwards so that the

counting and stroke systems engage again.

7. After the adjustment has been successfully completed, lock the adjustment

opening with a red CAL seal.

1. Insert side D of the pipette wrench

vertically into the lateral adjustment
opening of the pipette grip.

2. Move the pipette wrench into the

vertical position.

3. Turn the volume setting ring toward

– or +.
This adjusts the piston stoke of the
pipette. The digital display does not
change in the process.

Volume range

Vol. per revolution

0.1 – 2.5 μL

approx. 0.1 μL

0.5 - 10 μL

approx. 0.5 μL

2 - 20 μL

approx. 1 μL

10 - 100 μL

approx. 5 μL

20 - 200 μL

approx. 10 μL

100 - 1 000 μL

approx. 50 μL

500 - 5 000 μL

approx. 250 μL

1 - 10 mL

approx. 510 μL