Adjustment
Eppendorf SOP
English (EN)

1. Check whether the adjustment display is set

to "0".
If the adjustment display is not set to “0”, it
must be set to “0” first using the adjustment
tool. In this case, instead of continuing with
the factory adjustment changes, carry out a
gravimetric test of the Reference2 with the
adjustment display set to "0".

2. Use the plug to pierce the safety plug, and

remove it.

3. Insert the adjustment tool.

4. Hold the control button with one

hand.

1

1

ADJ

+

-

1000 µL

ADJ

R

e

ADJ

R

e

ADJ

R

e