2 adjusting adjustable volume pipettes, 1 reference variable, Adjusting adjustable volume pipettes 9.2.1 – Eppendorf SOP User Manual

Page 41: Reference variable

Advertising
background image

41

Technical specifications

Eppendorf SOP

English (EN)

9.2

Adjusting adjustable volume pipettes

9.2.1

Reference variable

Model

Test tip
epT.I.P.S. Color code
Volume range
Length

Testing
volume

Error limits

Error

Systematic error Random error

± %

± µL

± %

± µL

0.1 - 2.5 μL

Dark gray
0.1 - 10 μL
34 mm

0.25 μL

±12.0 ±0.03 ±6.0 ±0.015

1.25 μL

±2.5

±0.031 ±1.5 ±0.019

2.5 μL

±1.4

±0.035 ±0.7 ±0.018

0.5 - 10 μL

light gray
0.5 – 20 μL
46 mm

1 μL

±2.5

±0.025 ±1.8 ±0.018

5 μL

±1.5

±0.075 ±0.8 ±0.04

10 μL

±1.0

±0.1

±0.4 ±0.04

2 - 20 μL

light gray
0.5 – 20 μL
46 mm

2 μL

±3.0

±0.06 ±2.0 ±0.04

10 μL

±1.0

±0.1

±0.5 ±0.05

20 μL

±0.8

±0.16 ±0.3 ±0.06

2 - 20 μL

Yellow
2 - 200 μL
53 mm

2 μL

±5.0

±0.1

±1.5 ±0.03

10 μL

±1.2

±0.12 ±0.6 ±0.06

20 μL

±1.0

±0.2

±0.3 ±0.06

10 - 100 μL

Yellow
2 - 200 μL
53 mm

10 μL

±3.0

±0.3

±0.7 ±0.07

50 μL

±1.0

±0.5

±0.3 ±0.15

100 μL

±0.8

±0.8

±0.15 ±0.15

50 – 200 μL

Yellow
2 - 200 μL
53 mm

50 μL

±1.0

±0.5

±0.3 ±0.15

100 μL

±0.9

±0.9

±0.3 ±0.3

200 μL

±0.6

±1.2

±0.2 ±0.4

50 – 250 μL

Blue
50 - 1 000 μL
71 mm

50 μL

±1.4

±0.7

±0.3 ±0.15

100 μL

±1.1

±1.1

±0.3 ±0.3

250 μL

±0.6

±1.5

±0.2 ±0.5

100 – 1 000 μL Blue

50 – 1 000 μL
71 mm

100 μL

±3.0

±3.0

±0.3 ±0.3

500 μL

±1.0

±5.0

±0.2 ±1.0

1,000 μL ±0.6

±6.0

±0.2 ±2.0

500 – 2 500 μL Red

500 – 2 500 μL
115 mm

0.5 mL

±1.5

±7.5

±0.3 ±1.5

1.25 mL ±0.8

±10

±0.2 ±2.5

2.5 mL

±0.6

±15

±0.2 ±5.0

Advertising