3 multi-channel pipettes, 1 research, Multi-channel pipettes 9.3.1 – Eppendorf SOP User Manual

Page 49: Research

Advertising
background image

49

Technical specifications

Eppendorf SOP

English (EN)

9.3

Multi-channel pipettes

9.3.1

Research

Model

Test tip
epT.I.P.S. Color
code
Volume range
Length

Volume
in µL

Error limits

Error

Systematic

error

Random error

± %

± µL

± %

± µL

Research8-channel
0.5 - 10 μL

light gray
0.5 – 20 μL
46 mm

1

±8.0 ±0.08

±5.0 ±0.05

5

±4.0 ±0.2 ±2.0 ±0.1

10

±2.0 ±0.2 ±1.0 ±0.1

Research12-channel
0.5 - 10 μL

see 8-channel

Research8-channel
10 - 100 μL

Yellow
2 - 200 μL
53 mm

10

±3.0 ±0.3 ±2.0 ±0.2

50

±1.0 ±0.5 ±0.8 ±0.4

100 ±0.8 ±0.8 ±0.3 ±0.3

Research12-channel
10 - 100 μL

see 8-channel

Research8-channel
30 - 300 μL

Orange
20 – 300 μL
55 mm

30

±3.0 ±0.9 ±1.0 ±0.3

150 ±1.0 ±1.5 ±0.5 ±0.75

300 ±0.6 ±1.8 ±0.3 ±0.9

Research12-channel
30 - 300 μL

see 8-channel

Advertising