2 reference 2 fixed, Reference 2 fixed – Eppendorf SOP User Manual

Page 38

Advertising
background image

38

Technical specifications
Eppendorf SOP
English (EN)

9.1.2

Reference 2 fixed

Fixed-volume Reference2

Model

Test tip
epT.I.P.S. Color code
Volume range
Length

Eppendorf AG error limits

Error

Systematic error Random error

± %

± µL

± %

± µL

1 μL

Dark gray
0.1 μL – 10 μL
34 mm

±2.5

±0.025 ±1.8 ±0.018

2 μL

±2.0

±0.04

±1.2

±0.024

5 μL

Medium gray
0.1 μL – 20 μL
40 mm

±1.2

±0.06

±0.6

±0.03

10 μL

±1.0

±0.1

±0.5

±0.05

20 μL

light gray
0.5 μL – 20 μL L
46 mm

±0.8

±0.16

±0.3

±0.06

10 μL

Yellow
2 μL – 200 μL
53 mm

±1.2

±0.12

±0.6

±0.06

20 μL

±1.0

±0.2

±0.3

±0.06

25 μL

±1.0

±0.25

±0.3

±0.075

50 μL

±0.7

±0.35

±0.3

±0.15

100 μL

±0.6

±0.6

±0.2

±0.2

200 μL

±0.6

±1.2

±0.2

±0.4

200 μL

Blue
50 μL – 1 000 μL
71 mm

±0.6

±1.2

±0.2

±0.4

250 μL

±0.6

±1.5

±0.2

±0.5

500 μL

±0.6

±3.0

±0.2

±1.0

1,000 μL

±0.6

±6.0

±0.2

±2.0

Fixed-volume Reference2

Model

Test tip
epT.I.P.S. Color code
Volume range
Length

Eppendorf AG error limits

Error

Systematic error Random error

± %

± mL

± %

± mL

2.0 mL

Red
0.5 mL – 2.5 mL
115 mm

±0.6

±0.012

±0.2

±0,004

2.5 mL

±0.6

±0.015

±0.2

±0.005

Advertising