13

Sterilization and cleaning

Eppendorf SOP

English (EN)

4.1.3

Research and Research pro

With the Research and Research pro pipettes, the lower section is autoclavable.

With single-channel models prior to autoclaving

1. Pull off the ejector sleeve with depressed ejector.
2. Unscrew the lower part of the pipette.

With the Multi-channel versions, you can autoclave the entire lower section.

After autoclaving, assemble the parts only after room temperature has been
reached and the parts have been dried.

4.1.4

Top Buret

The Top Buret can not be autoclaved.

4.1.5

Varipette and Multipette

The

• Varipette,
• Multipette,
• Multipetteplus,
• Multipette stream
• MultipetteXstream
• MultipetteM4

are not autoclavable.

4.1.6

Varispenser

The Varispenser and Varispenser plus are only autoclavable when fully
assembled.

With theVarispenser plus

1. Set the discharge valve toggle to the dispensing position (–>).
2. Release the volume setting knob, move to the middle position and leave

released.

3. Place the dispenser on a cloth and autoclave, avoiding contact with hot

surfaces.

4. Only use the dispenser again when it has cooled down to room temperature.