3 research fixed, Research fixed – Eppendorf SOP User Manual

Page 39

Advertising
background image

39

Technical specifications

Eppendorf SOP

English (EN)

9.1.3

Research fixed

Model

Test tip
epT.I.P.S. Color code
Volume range
Length

Error limits

Error

Systematic error

Random error

± %

± µL

± %

± µL

10 μL

Yellow
2 - 200 μL
53 mm

±1.2

±0.12

±0.6

±0.06

20 μL

±1.0

±0.2

±0.3

±0.06

25 μL

±1.0

±0.25

±0.3

±0.075

50 μL

±0.7

±0.35

±0.3

±0.15

100 μL

±0.6

±0.6

±0.2

±0.2

200 μL

Blue
0.05 - 1 mL
71 mm

±0.6

±1.2

±0.2

±0.4

250 μL

±0.6

±1.5

±0.2

±0.5

500 μL

±0.6

±3.0

±0.2

±1.0

1,000 μL

±0.6

±6.0

±0.2

±2.0

Advertising