Eppendorf SOP User Manual

Page 29

Advertising
background image

29

Adjustment

Eppendorf SOP

English (EN)

6. This results in the following approximate volume changes:

The values mentioned are theoretical values and are for orientation purposes
only. The volume changes mentioned apply to each volume setting.

single-channel
Nominal volume
Color code

+½ 
revolution

+¼ 
revolution

–¼ 
revolution

–½ 
revolution

2.5 μL
Dark gray

0.106 μL

0.053 μL

-0.053 μL

-0.106 μL

10 μL
Medium gray

0.53 μL

0.27 μL

-0.27 μL

-0.53 μL

20 μL
light gray

1.06 μL

0.53 μL

-0.53 μL

-1.06 μL

20 μL
Yellow

1.07 μL

0.54 μL

-0.54 μL

-1.07 μL

100 μL
Yellow

5.4 μL

2.7 μL

-2.7 μL

-5.4 μL

200 μL
Yellow

10.8 μL

5.4 μL

-5.4 μL

-10.8 μL

300 μL
Orange

10.7 μL

5.4 μL

-5.4 μL

-10.7 μL

1,000 μL
Blue

54 μL

27 μL

-27 μL

-54 μL

5 mL
Purple

271 μL

135 μL

-135 μL

-271 μL

10 mL
Turquoise

542 μL

271 μL

-271 μL

–542 μL

multi-channel
Nominal volume
Color code


revolution

+¼ 
revolution

–¼ 
revolution

–½
revolution

10 μL
Medium gray

0.53 μL

0.27 μL

-0.27 μL

-0.53 μL

100 μL
Yellow

5.4 μL

2.7 μL

-2.7 μL

-5.4 μL

300 μL
Orange

10.7 μL

5.4 μL

-5.4 μL

-10.7 μL

Advertising