Technical specifications

Eppendorf SOP

English (EN)

9.3.3

Research plus

Research plus adjustable multi-channel

Model

Test tip
epT.I.P.S. Color code
Volume range
Length

Testing
volume

Eppendorf AG error limits

Error

Systematic

error

Random

error

± %

± µL

± %

± µL

0.5 μL – 10 μL
Increment:
0.01 μL

Medium gray
0.1 μL – 20 μL
40 mm

0.5 μL ±12

±0.06 ±8.0

±0.04

1 μL

±8.0 ±0.08 ±5.0

±0.05

5 μL

±4.0 ±0.2

±2.0

±0.1

10 μL

±2.0 ±0.2

±1.0

±0.1

10 μL – 100 μL
Increment:
0.1 μL

Yellow
2 μL - 200 μL
53 mm

10 μL

±3.0 ±0.3

±2.0

±0.2

50 μL

±1.0 ±0.5

±0.8

±0.4

100 μL ±0.8 ±0.8

±0.3

±0.3

30 μL – 300 μL
Increment:
0.2 μL

Orange
20 μL – 300 μL
55 mm

30 μL

±3.0 ±0.9

±1.0

±0.3

150 μL ±1.0 ±1.5

±0.5

±0.75

300 μL ±0.6 ±1.8

±0.3

±0.9