2 biomaster, 3 reference variable, Biomaster – Eppendorf SOP User Manual

Page 19: Reference variable

Advertising
background image

19

Adjustment

Eppendorf SOP

English (EN)

7.2.2

Biomaster

Auxiliary equipment
• Supplied pipette wrench (order no. 4910 092.001)

7.2.3

Reference variable

Auxiliary equipment
• Supplied pipette wrench (order no. 4910 092.001)
• Supplied red CAL seal

1. Use the pipette wrench to adjust the

volume display of the pipette, with
the piston stroke unchanged, to the
actual volume value of the
measurement.

2. Pull off the pipette wrench.
3. Set the pipette to the target volume

in the usual manner.

4. Mark the adjustment on the pipette.

1. Insert side B of the pipette wrench

into the adjustment opening in the
lid.

2. Using the pipette wrench, set the

digital display of the pipette, with
unaltered piston stroke, to the actual
volume value of the measurement.

3. Pull off the pipette wrench.
4. Set the pipette to the target volume

in the usual manner.

5. After the adjustment has been

successfully completed, lock the
adjustment opening with a red CAL
seal.

10

20

06

26