4 single-stroke dispenser (varispenser), 5 piston stroke burets, Single-stroke dispenser (varispenser) – Eppendorf SOP User Manual

Page 60: Piston stroke burets

Advertising
background image

60

Technical specifications
Eppendorf SOP
English (EN)

9.6.4

Single-stroke dispenser (Varispenser)

9.6.5

Piston stroke burets

Nominal volume

Error limits

Measurement errors

Systematic

Random

± %

± µL

± %

± µL

0.01 ml

±2.0 %

±0.2 μl

±1.0 %

±0.1 μl

0.02 ml

±2.0 %

±0.4 μl

±0.5 %

±0.1 μl

0.05 ml

±1.5 %

±0.75 μl

±0.4 %

±0.2 μl

0.1 ml

±1.5 %

±1.5 μl

±0.3 %

±0.3 μl

0.2 ml

±1.0 %

±2.0 μL

±0.3 %

±0.6 μl

0.5 ml

±1.0 %

±5.0 μL

±0.2 %

±1.0 μL

1 ml

±0.6 %

±6.0 μL

±0.2 %

±2.0 μL

2 ml

±0.6 %

±12.0 μL

±0.2 %

±4.0 μL

5 mL

±0.6 %

±30.0 μL

±0.2 %

±10.0 μL

10 mL

±0.6 %

±60.0 μL

±0.2 %

±20.0 μL

25 ml

±0.6 %

±150.0 μL

±0.2 %

±50.0 μL

50 mL

±0.6 %

±300.0 μL

±0.2 %

±100.0 μL

100 ml

±0.6 %

±600.0 μL

±0.2 %

±200.0 μL

200 ml

±0.6 %

±1 200 μL

±0.2 %

±400,0 μL

Nominal volume

Error limits

Measurement errors

Systematic

Random

± %

± µL

± %

± µL

<1 mL

±0.6 %

±6,0 μL

±0.1 %

±1,0 μL

2 ml

±0.5 %

±10,0 μL

±0.1 %

±2,0 μL

5 mL

±0.3 %

±15,0 μL

±0.1 %

±5,0 μL

10 mL

±0.3 %

±30,0 μL

±0.1 %

±10,0 μL

20 ml

±0.2 %

±40,0 μL

±0.1 %

±20,0 μL

25 ml

±0.2 %

±50,0 μL

±0.1 %

±25,0 μL

50 mL

±0.2 %

±100,0 μL

±0.1 %

±50,0 μL

100 ml

±0.2 %

±200 μL

±0.1 %

±100,0 μL