3 research adjustable-volume, Research adjustable-volume – Eppendorf SOP User Manual

Page 44

Advertising
background image

44

Technical specifications
Eppendorf SOP
English (EN)

9.2.3

Research adjustable-volume

Model

Test tip
epT.I.P.S. Color code
Volume range
Length

Testing
volume

Error limits

Error

Systematic

error

Random

error

± %

± µL

± %

± µL

0.1 - 2.5 μL Dark gray

0.1 - 10 μL
34 mm

0.25

μL

±12.0 ±0.03 ±6.0 ±0.015

1.25 μL

±2.5

±0.031 ±1.5

±0.019

2.5 μL

±1.4

±0.035 ±0.7

±0.018

0.5 - 10 μL

light gray
0.5 – 20 μL L
46 mm

1 μL

±2.5

±0.025 ±1.8

±0.018

5 μL

±1.5

±0.075 ±0.8

±0.04

10

μL

±1.0 ±0.1 ±0.4 ±0.04

2 - 20 μL

Yellow
2 - 200 μL
53 mm

2

μL

±5.0 ±0.1 ±1.5 ±0.03

10 μL

±1.2

±0.12

±0.6

±0.06

20

μL

±1.0 ±0.2 ±0.3 ±0.06

10 - 100 μL Yellow

2 – 200 μL
53 mm

10

μL

±3.0 ±0.3 ±1.0 ±0.1

50

μL

±1.0 ±0.5 ±0.3 ±0.15

100

μL

±0.8 ±0.8 ±0.2 ±0.20

20 - 200 μL Yellow

2 - 200 μL
53 mm

20

μL

±2.5 ±0.5 ±0.7 ±0.14

100

μL

±1.0 ±1.0 ±0.3 ±0.3

200

μL

±0.6 ±1.2 ±0.2 ±0.4

100 –
1 000 μL

Blue
0.05 – 1 mL
71 mm

100

μL

±3.0 ±3.0 ±0.6 ±0.6

500

μL ±1.0 ±5.0 ±0.2 ±1.0

1,000

μL

±0.6 ±6.0 ±0.2 ±2.0

0.5 – 5 mL

Purple
0.1 – 5 mL
120 mm

0.5 mL

±2.4

±12

±0.6

±3.0

2.5 mL

±1.2

±30

±0.25

±6.25

5.0 mL

±0.6

±30

±0.15

±7.5

1 – 10 mL

Turquoise
1 – 10 mL
165 mm

1.0 mL

±3.0

±30

±0.6

±6.0

5.0 mL

±0.8

±40

±0.2

±10

10.0 mL

±0.6

±60

±0.15

±15

Advertising