Technical specifications

Eppendorf SOP

English (EN)

9

Technical specifications

The following specifications are valid for the following conditions:

9.1

Fixed-volume pipettes

9.1.1

Reference fixed

Liquid:

Distilled or deionized water

Reference temperature: 20 °C to 25 °C ±0.5 °C

Number of
determinations:

10, in accordance with EN ISO 8655 with original
Eppendorf pipette tips

Technical specifications subject to change!

Model

Test tip
epT.I.P.S. Color code
Volume range
Length

Error limits

Error

Systematic error

Random error

± %

± µL

± %

± µL

1 μL

light gray
0.5 - 20 μL L
46 mm

±2.5

±0.025 ±1.8

±0.018

2 μL

±2.0

±0.04

±1.2

±0.024

5 μL

±1.5

±0.075

±0.8

±0.04

10 μL

±1.0

±0.1

±0.5

±0.05

10 μL

Yellow
2 - 200 μL
53 mm

±1.0 ±0.1 ±0.5

±0.05

20 μL

±0.8

±0.16

±0.3

±0.06

25 μL

±0.8

±0.2

±0.3

±0.075

50 μL

±0.7

±0.35

±0.3

±0.15

100 μL

±0.6

±0.6

±0.2

±0.2

200 μL

Blue
50 - 1 000 μL
71 mm

±0.6

±1.2

±0.2

±0.4

250 μL

±0.6

±1.5

±0.2

±0.5

500 μL

±0.6

±3.0

±0.2

±1.0

1,000 μL

±0.6

±6.0

±0.2

±2.0

1 500 μL

Red
500 – 2 500 μL
115 mm

±0.6

±9.0

±0.2 ±3.0

2 000 μL

±0.6

±12

±0.2

±4.0

2,500 μL

±0.6

±15

±0.2

±5.0

Advertising