Operating the Model 505.60 HPU Locally or Remotely

Model 505.60 - 505.180 SilentFlo™ HPU

Operation

75

4. Verify that at least one pump is enabled. Enable additional pumps as

necessary.

5. Press Run/Low Pressure to start the pumps. The pumps start sequentially

to prevent an electrical overload. Each pump starts in low pressure mode.

6. Check the HPU for leaks and unusual sounds. Stop the HPU immediately if

leaks or unusual sounds are noted. Determine the cause and fix the problem
before restarting the HPU.

7. Press High Pressure to turn on high hydraulic pressure.

Note

If the HPU generates an interlock during operation (such as low fluid
level or high temperature), the HPU will stop. Once the cause has been
corrected, press the Reset button before restarting.

8. Run the HPU for about 30 minutes or until the hydraulic fluid is up to

operating temperature [typically

43–49°C (110–120°F)

] before using your

test system.

9. When in high pressure mode, press the High Pressure button to return to

low pressure mode.

10. To stop the HPU, press the Stop button.

Remote operation

Remote operation can occur at a remote control panel or from your system
controller. The controller must be connected to the HPU.

1. Make a general inspection of the HPU. Ensure that all cooling water valves

are open. Ensure the Emergency Stop button is released.

2. If not already displayed, press the Main Display button to open the Main

Display screen.

3. Verify that the Remote button is green. If the button is not green, press the

button to change it to green (indicating remote mode).

4. Verify that at least one pump is enabled. Enable additional pump as

necessary.

5. Start the HPU in low pressure at your controller.

6. Check the HPU for leaks and unusual sounds. Stop the HPU immediately if

leaks or unusual sounds are noted. Determine the cause and fix the problem
before restarting the HPU.

7. Select high pressure at your system controller.

Note

If the HPU generates an interlock during operation (low fluid level or high
temperature), the HPU will be stopped. Once the cause has been
corrected, press the Reset button on the HPU user interface before
restarting.

8. Run the HPU for about 30 minutes or until the hydraulic fluid is up to

operating temperature [typically 43–49°C (110–120°F)] before using your
test system.

9. Stop the HPU with the controller.