MTS Series 505 SilentFlo Hydraulic Power Unit Model 505-180 User Manual

Page 82

Advertising
background image

Model 505.60 - 505.180 SilentFlo™ HPU

82

Adjusting the Model 505.60 Hydraulic Pressure

Operation

1. If not already selected, press the Main Display button on the user interface

panel to select the Main Display.

2. Press the Pump #_ Enabled for the first pump to be adjusted.

Press Pump #_ Disabled for all the other pumps.

3. Press Run/Low Pressure to start the HPU in low pressure. Ensure that there

are no flow demands by the system. Run the pump in this mode until the
hydraulic fluid is at normal operating temperature.

4. Press the High Pressure button to select high pressure mode.

5. Loosen the nut securing the output pressure control.

6. While an assistant monitors the Hydraulic Fluid Pressure on the Main

Display screen, adjust the output pressure as follows until the desired
pressure is displayed.

Turn the output pressure control clockwise to increase the pressure.

Turn the output pressure control counterclockwise to decrease the
pressure.

7. Hold the output pressure control to prevent it from moving and tighten the

nut to secure it.

8. Check the Hydraulic Fluid Pressure on the Main Display screen to ensure

that the desired hydraulic pressure is being maintained.

9. Press the Pump #_ Enabled for the next pump to be adjusted.

Press Pump #_ Disabled for the previously adjusted pump.

10. Repeat Steps 5 through 9 until all pumps have been adjusted to the desired

pressure.

11. After the output pressure has been adjusted on all the pumps, enable the

pumps you want operational for testing.