Sensaphone IMS-4000 Users manual User Manual

Page 22

Advertising
background image

IMS-4000 Manual

20

on/off swiTCh

The on/off switch connects main power and battery power to the system. Note that the system will only turn on when main

power is present. When main power fails, the internal battery will automatically supply power to the unit.

Figure 2: Rear Panel of the IMS Host

insTallaTion

This section provides information on:

• Operating environment
• Rack, wall, and tabletop installation
• Connecting sensors

parts required

Phillips Screwdriver • 9 pin F/F null modem cable • Dumb terminal or PC

w/9 pin com port

operaTinG enVironMenT

Before you install the IMS-4000 Host be sure that your operating environment meets the physical requirements of the

equipment.

Operating Temperature: 32º–122º Fahrenheit (0º–50º C)

Humidity: 5–90 %RH, non-condensing

Power: 100–250VAC 50–60 Hz outlet within 6'

Rack Requirements: Standard 19" equipment rack with supplied mounting bracket hardware.

Requires 1.0 EIA rack mount space.

Tabletop requirements: Flat area which can support an enclosure 17" wide by 10" deep by 2" high.

fUse replaCeMenT

The IMS-4000 host contains a user replaceable fuse on the main AC power supply, as well as a spare fuse. Both are built into

the power connector and can be accessed with a small flat head screwdriver. See figure 2. Before attempting to replace the

fuse, turn the power switch off and disconnect the power cable. After you remove the cord you will see a small notch along

the bottom edge of the cable entry area of the connector. Insert a small flat-head screwdriver into the notch and force the

fuse holder away from the connector body. There will be a visible fuse held in place by a plastic clamp - this is the fuse that

is actively protecting the device. There will also be a spare fuse located inside the square plastic cavity below the notch. Use

the screwdriver to push out the spare fuse. Replace the blown fuse with the spare and insert the fuse holder back into the

power entry connector. Reattach the power cord and turn the power switch on.

Replacement Fuse: 1.6Amp 250VAC Fast acting 5X20 mm (e.g. Littelfuse # 023501.6HXP)

Advertising