Sunnex PF Series MRI Service & Installation Manual

10

DOES NOT VOID THE WARRANTY OF THE PRODUCT AND DOES
NOT VOID THE CERTIFICATION LISTING). Install (solder) wires of
other end of lamp cord to the selected D sub connector for inside of the
MRI room and install this connector to the RF filter from inside the MRI
room. (Sunnex Inc. does not provide RF filters or the mating D sub
connectors to RF filter). Sunnex Inc. can provide the output cord
(Sunnex part# 50710030) or 18AWG/2 wire, temperature rating 105
degree C should be used. A typical installation is shown in following
diagram. A 3-pin, 9-pin, 25-pin D Sub connector can be used.


5. Plug male connector (from lamp cord) into the female connector on the

electrical box. CAUTION: Male connector has a mechanical latch. Be
sure lamp connector is fully engaged into electrical box female connector
and the latch is fully engaged. (To disconnect, press tab on the female
connector [and hold] and pull the male connector.)


Figure 2. Lamp connection with electrical box

Lamp head

connector

Electrical

box

RF FILTER

D-SUB

CONNECTOR

D-SUB

CONNECTOR

SUNNEX

POWER

SUPPLY

BOX

SUNNEX

PF MRI

LIGHT

RF PANEL OF

MRI ROOM