Sunnex PF Series MRI Service & Installation Manual

4

2. INTRODUCTION

The objective of this manual is to assist you in assembling your PF series
MRI light, to introduce some of its most important features, and to provide
suggestions for its care.

2.1 INTENDED USE

The Sunnex PF series MRI light is intended for use in MRI environments
where illumination is required for various tasks around the bore.

CAUTION!

This light is intended for out of bore procedures only. Any deviation or
inappropriate use may result in serious injury.


2.2 SAFETY INSTRUCTIONS

The Sunnex PF series MRI light must be operated with the safety of the user
and patient in mind. Please read this manual and any other instructions
before installing the equipment. Please make sure that all appropriate service
staff is informed and educated on the equipment to ensure continued safety
and needed maintenance. Be sure you understand the intended use of this
equipment and recommended operation before attempting to install and use
the equipment.

IMPORTANT!

This light is MRI conditional and can be used as a medical task

light around the bore. NOTE! It is extremely important that all
guidelines are strictly followed to ensure safety for users and
patients.

Use the light only for its intended purpose. It is a portable medical

task light intended for use outside of the bore.

WARNING! The power supply contains ferrous materials. DO NOT

bring the power supply close to the magnet (100 Gauss line).
Always take the furthest route away from the magnet when
moving the power supply in the MRI room.

The power supply must be fastened to a fixed structure.

ALWAYS bring the light outside the MRI room when performing

any service on the light. This includes any cleaning and/or bulb
replacements. Also, assemble the light outside of the MRI room.