Sunnex PF Series (MRI) User Manual

Page 25

Advertising
background image

Sunnex PF Series MRI Service & Installation Manual

25

Guidance and Manufacturer’s Declaration Emissions Equipment

and Systems that are NOT Life-Supporting

The PF MRI Light (PF 28016-M) is intended for use in the electromagnetic
environment specified below. The customer or user of the PF MRI Light
(PF 28016-M) should ensure that it is used in such an environment.

Immunity

Test

IEC

60601

Test

Level

Complia

nce

Level

Electromagnetic Environment

Guidance

Conducted
RF
IEC 61 000-
4-6

3 Vrms
from

150 kHz
to
80 MHz

V1 = 3
Vrms

Portable and mobile RF
communications equipment should
be separated from the PF MRI Light
(PF 28016-M) by no less than the
recommended separation distances
calculated/listed below:
D = (3.5/V1)

P

Radiated
RF
IEC 61000-
4-3

3 V/m

80 MHz
to
2.5 GHz

E1 =
3V/m

D = (3.5/ E1)/

P

80 to 800 MHz

D = (7 / E1)/

P

800 MHz to 2.5 GHz


Where P is the maximum power
rating in watts and D is the
recommended separation distance in
meters.

Field strengths from fixed
transmitters, as determined by an
electromagnetic site survey, should
be less that the compliance levels
(V1 and E1).

Interference may occur in the vicinity
of equipment containing a transmitter.