Sunnex PF Series MRI Service & Installation Manual

7

This light is packaged in card board and should be transported protected
from moisture. The light should be stored in a dry environment and room
temperature.

Transportation and storage temperature: -5°F / -20°C to 140°F / 60°C

Humidity < 95%

4. ASSEMBLY INSTRUCTIONS

There are several factors involved with proper installation. Considering them
will help you produce a safe work environment for yourself and your patients.

Please follow the directions for assembly carefully as any negligence or
poorly performed assembly may void the Warranty. Any modification to the
structure of the light may also void the listings of this product.

4.1 PF SERIES MRI MOBILE - PF28016-M


CAUTION! THE ELECTRICAL BOX CONTAINS FERROUS
MATERIALS AND MUST BE HANDLED WITH CAUTION.
THE ELECTRICAL BOX SHOULD BE SECURED TO A
STRUCTURAL MEMBER AT A SAFE DISTANCE (100
GAUSS LINE) AWAY FROM THE MRI MACHINERY. DO
NOT MOUNT BOX DIRECTLY TO SHEET ROCK OR
PLYWOOD SUBSTRATES. IDEALLY, THE ELECTRICAL
BOX IS SECURED BY NON-FERROUS FASTENERS OR
VELCRO, DIRECTLY TO THE FLOOR OUTSIDE THE
MAXIMUM RECOMMENDED GAUSS LINE.


It is recommended the electrical box is mounted away from high traffic area
and congestion to avoid the possibility of damage or removal from mounting
surface.

Assembly and installation of the light

IMPORTANT! Assemble the light outside of the MRI room.