Sunnex PF Series MRI Service & Installation Manual

21

9. TROUBLESHOOTING

Please refer to the following table for troubleshooting.



CAUTION! DISCONNECT THE LIGHT FROM THE POWER
SUPPLY BEFORE ATTEMPTING ANY OF THE
ELECTRICAL CHECKS MENTIONED BELOW.


PROBLEM POSSIBLE

CAUSE

SOLUTION

No light from light
head

Light head is on OFF
position

Turn on the ON/OFF switch
to ON position

No light from light
head, when ON/OFF
switch is ON

Bulb has not been
installed or it has
burned out

Refer to bulb replacement
procedure in the manual

Fuse is blown

Inspect / replace the fuse
following the fuse
replacement procedure

No power to the
electrical box

Check for facility power to the
electrical box

Power Supply unit is
malfunctioning

Call Sunnex Customer
Service

Incorrect

bulb

installed

Install correct bulb

Bulb burns out
quickly

Incorrect bulb
installed in the light
head

Install correct bulb

Light output is
irregular in its shape
or intensity

Incorrect bulb
installed in the light
head

Install correct bulb

Bulb/reflector is not
seated in its holder
correctly

Adjust the bulb/reflector so it
seats in the holder properly.

No light from light

Circuit or wiring

Check all wiring connections.