Sunnex PF Series MRI Service & Installation Manual

26

Recommended Separation Distances Between Portable and

Mobile RF Communications Equipment and the PF MRI Light (PF

28016-M) Equipment and Systems that are NOT Life-Supporting

The PF MRI Light (PF 28016-M) is intended for use in the
electromagnetic environment in which radiated disturbances are
controlled. The customer or user of the PF MRI Light (PF 28016-M)
can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF Communications
Equipment and the PF MRI Light (PF 28016-M) as recommended
below, according to the maximum output power of the
communications equipment.

Recommended Separation Distances for the PF

MRI Light (PF 28016-M)

Maximum

Output Power

(Watts)

150 kHz to 80

MHz

d = 1.1667

P

80 MHz to

800MHz

d = 1.1667

P

800 MHz to 2.5

GHz

d = 2.3333

P

0.01 0.11667 0.11667 0.23333

0.1

0.36894 0.36894 0.73785

1

1.1667 1.1667 2.3333

10

3.6894 3.6894 7.3785

100 11.667 11.667 23.333