Sunnex PF Series MRI Service & Installation Manual

19

On electrical box

CAUTION!

THIS POWER SUPPLY IS MAGNETIC

THIS POWER SUPPLY MUST BE KEPT BEYOND 100

GAUSS LINE OF MRI MAGNET

CE TRANSFORMATEUR EST FERRO-MAGNÉTIQUE.

CE TRANSFORMATEUR, DOIT SE TROUVER EN DEHORS

DE LIGNES DES 100 GAUSS EST DU CHAMPS

MAGNÉTIQUE DE D’IRM


For 120V

Sunnex CE INPUT 120V

MODEL PF 28016-M Hz 60
S/N OUTPUT 14V

AMPS 1.5
WATTS 50W
• FUSES LOCATED ON TRANSFORMER BOX

T-2A-250V

2 AMP, ¼” X 1 ¼”

• POWER CORD IS SPECIFIED FOR USE WITH HOSPITAL GRADE RECEPTACLE

• REMOVE POWER CORD FROM WALL RECEPTACLE BEFORE SERVICING
• FUSIBLES PLACEES SOUS COUVERTURES

T-2A-250V

2 AMP, ¼” X 1 ¼”

• LE CABLE D’ALIMENTATION EST SPECIFIQUE A UNE UTILISATION EN MILIEU HOSPITALIER

• DEBRANCHER LE CABLE D’ALIMENTATION AVANT TOUT DEMONTAGE DE LA LAMPE

SUN-6


For 230V

Sunnex CE INPUT 230V

MODEL PF 28016-M Hz 50
S/N OUTPUT 14V

AMPS 1.0
WATTS 50W
• FUSES LOCATED ON TRANSFORMER BOX

T-1.25A-250V

1.25 AMP, ¼” X 1 ¼”

• POWER CORD IS SPECIFIED FOR USE WITH HOSPITAL GRADE RECEPTACLE

• REMOVE POWER CORD FROM WALL RECEPTACLE BEFORE SERVICING
• FUSIBLES PLACEES SOUS COUVERTURES

T-1.25A-250V

1.25 AMP, ¼” X 1 ¼”

• LE CABLE D’ALIMENTATION EST SPECIFIQUE A UNE UTILISATION EN MILIEU HOSPITALIER

• DEBRANCHER LE CABLE D’ALIMENTATION AVANT TOUT DEMONTAGE DE LA LAMPE