3 mounting the robotic arm reinforcement, Mounting the robotic arm reinforcement, Figure 12 – Metrohm 815 Robotic USB Sample Processor XL User Manual

Page 29: Mounting the 6.2058.090 robotic arm reinforcement

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

815 Robotic USB Sample Processor XL

■■■■■■■■

21

3.4.3

Mounting the robotic arm reinforcement

In cases in which it is necessary to take up or strip off a filter or a pipetting
tip on the robotic arm, you must mount a 6.2058.090 robotic arm
reinforcement
. The reinforcement provides the robotic arm with stable
resistance when picking up or stripping the tools used and prevents it
from bending.

The reinforcement has to be placed before mounting a robotic arm. Pro-
ceed as follows:

2

1

6.2058.090

Figure 12

Mounting the 6.2058.090 robotic arm reinforcement

1 Shift the reinforcement from the front over the holder plate of the

Swing Head.

2 Screw the reinforcement to the lift guide with the screws provided.

Advertising