12 connecting an external pump, Figure 25, Connecting the pump – Metrohm 815 Robotic USB Sample Processor XL User Manual

Page 42

Advertising
background image

3.12 Connecting an external pump

■■■■■■■■■■■■■■■■■■■■■■

34

■■■■■■■■

815 Robotic USB Sample Processor XL

3.12

Connecting an external pump

If no integrated pump or a Sample Processor model version without
pumps is used, up to two external pumps per tower can be connected.

The 843 Pump Station (as model version with membrane pumps or with
peristaltic pumps) has two pump drives and is connected to two sockets
of the Sample Processor via the 6.2141.300 connecting cable (double
cable with two plugs). The 772 Pump Unit (peristaltic pump) and the
823 Membrane Pump Unit (membrane pump) have a firmly mounted
connection cable with a single plug.

Connecting the pump

Figure 25

Connecting the pump

Connect an external pump as follows:

1

Plug the threaded plug of the connection cable into one of the
connection sockets Ext. pump 1 or Ext. pump 2 on the rear of
a tower of the Sample Processor.
Correct alignment of the 3 contact pins must be observed.

Tighten the knurled screw at the front end of the plug by hand in
clockwise direction. This will secure the plug.

2 In case of an 843 Pump Station, connect the other end of the cable

(9-pin D-Sub plug) to the socket Remote 1 of the pump.

Advertising