5 Troubleshooting
815 Robotic USB Sample Processor XL
■■■■■■■■
47
Problem
Cause
Remedy
Swing Head – The Swing
Head drive is defective.
Contact the Metrohm Service.
5.3
Pump
Problem
Cause
Remedy
The pump is leaking.
Sample Processor – A tub-
ing connection is leaking.
Check the tubing connections especially
between the distributor an the pump and seal
tightly.
Canister – There is too
much pressure on the
pump valve.
■
Ensure that the canisters are not placed on
a higher level than the pump.
■
Check the fill level of the canisters.