5 Troubleshooting

815 Robotic USB Sample Processor XL

■■■■■■■■

47

Problem

Cause

Remedy

Swing Head – The Swing
Head drive is defective.

Contact the Metrohm Service.

5.3

Pump

Problem

Cause

Remedy

The pump is leaking.

Sample Processor – A tub-
ing connection is leaking.

Check the tubing connections especially
between the distributor an the pump and seal
tightly.

Canister – There is too
much pressure on the
pump valve.

Ensure that the canisters are not placed on
a higher level than the pump.

Check the fill level of the canisters.