Figure 16, Mounting the robotic arm – Metrohm 815 Robotic USB Sample Processor XL User Manual

Page 33

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

815 Robotic USB Sample Processor XL

■■■■■■■■

25

After the initialization of the Sample Processor, the drive disc of the Swing
Head is positioned as though the robotic arm were located in the outer-
most position.

1

2

3

Figure 16

Mounting the robotic arm

Mount the robotic arm as follows:

1 Place the 6.1458.040 titration head insert in the opening of the

robotic arm and screw tight with the supplied screws.

2 Hold the robotic arm in such a way that the opening faces to the

right. While doing so, rotate the robotic arm outwards as far as pos-
sible, i.e. towards the tower - see above. Slip the robotic arm from
below over the guide pins of the drive disc of the Swing Head.

Note

Take care to ensure that you do not twist the drive disc, thus caus-
ing pressure against the drive.

3 Screw the robotic arm to the Swing Head tightly with the screws and

washers provided.