3.9 Guide chain for cables and tubing

■■■■■■■■■■■■■■■■■■■■■■

30

■■■■■■■■

815 Robotic USB Sample Processor XL

Figure 20

Guide chain - Opening chain links

2 Insert into the guide chain

Place the required tubings or cables into the guide chain.

3 Close the guide chain

Close the clip for each chain link again by hand and apply forceful
pressure to snap them into place.

Caution

Take care to ensure when mounting tubing and cables that there is no
traction on the drives while moving the lift or swiveling the robotic arm.
This could lead to overloading of and possible damage to the drive.

Remove the clips of the two lowest chain links when you install the
rinsing and aspiration tubing.