2.1 Setting up the instrument

818 IC Pump / 8.818.1023 Instructions for Use

7

2 Installation

2.1

Setting up the instrument

2.1.1 Packaging

The 818 IC Pump is supplied together with the separately packed

accessories in special packagings containing shock-absorbing foam

linings designed to provide excellent protection. The actual instrument

is packed in an evacuated polyethylene bag to prevent the ingress of

dust. Please store all these special packagings as only they assure

transport of the instruments free from damage.

2.1.2 Check

After receipt, immediately check whether the shipment is complete and

has arrived without damage (compare with delivery note and list of

accessories in section 6.2). In the case of transport damage, see

instructions in section 6.5.1 "Warranty".

2.1.3 Location

Position the 818 IC Pump in the laboratory at a location convenient for

operation, free from vibrations and protected against a corrosive

atmosphere and contamination by chemicals.

To avoid disturbing temperature influences on the insulated column

compartment, the pump and eluent reservoir must be protected

against direct sunlight.

2.1.4

Arrangement of the instruments

The 818 IC Pump should always be at the very bottom of a IC system.

The 818 IC Pump should always be placed at the bottom of the

instrument stack so that any leaks which may occur in the pump

tubing or connections cannot cause damage to the other instruments

by leakage liquids.