6 mains connection, Mains connection, Setting the mains voltage – Metrohm 818 IC Pump User Manual

Page 21: Ction 2.6, Ction 2.6.1

Advertising
background image

2.6 Mains connection

818 IC Pump / 8.818.1023 Instructions for Use

15

2.6 Mains

connection

Follow the instructions below for connecting to the power supply. If

the instrument is operated with a mains voltage set wrongly and/or

wrong mains fuse, there is a danger of fire!

2.6.1

Setting the mains voltage

Before switching on the 818 IC Pump for the first time, check whether

the mains voltage set on the instrument (can be read in the mains

voltage selector 11) matches the local mains voltage. If this is not the

case, you must switch the mains voltage by pushing the mains voltage

selector 11 with a screwdriver:



1 Disconnect mains cable

Disconnect mains cable from mains connection plug 10 of the

818 IC Pump.

2 Remove fuse cover

Using a screwdriver, lever out fuse cover 10 forwards until it

opens.

3 Check and change fuses if necessary

Carefully take the fuses installed for the desired mains voltage

out of the fuse holder and check its specifications:

100

120 V 0.5 A (slow blow) Metrohm No. U.600.0013

220

240 V 0.25 A (slow blow) Metrohm No. U.600.0010

Change fuse if necessary and reinsert in fuse holder.

4 Install fuse holder

Reinsert fuse holders in the instrument (the arrows printed on the

holders must point in the same direction as the arrows on the

inside of fuse cover 10).

5 Install fuse cover

Push in fuse cover 10 firmly until it clicks into place.

6 Connect mains cable

Plug mains cable into mains connection plug 10.


Position of the mains
voltage selector 11:

230V: 220

240 V

±

10%

115V: 110

120 V

±

10%

230V

115V