3 faults and malfunctions, Faults and malfunctions – Metrohm 818 IC Pump User Manual

Page 36

Advertising
background image

4 Notes – Maintenance – Faults

818 IC Pump / 8.818.1023 Instructions for Use

30

4.3 Faults

and

malfunctions

If difficulties appear with the IC system during analyses, their causes

are best investigated in the order separating column

818 IC Pump

eluent

819/820 IC System (see 819/820 Instructions for Use,

section Faults and malfunctions). Several of the malfunctions which may

appear with the 818 IC Pump are listed in the following table with details

of possible causes and countermeasures.

Malfunction

Cause

Rectification

Baseline with high
noise level, pulsation

Contaminated pump
values

Faulty piston seals

Quality of the pump
does not suffice for the
selected sensitivity

Clean the valves (see
section 4.2.2)

Replace the piston
seals (see
section 4.2.2)

Use pulsation
dampener, use more
powerful pump or
lower the sensitivity

Drift of baseline

Thermal equilibrium
not yet reached

Leak in system

Evaporation of organic
solvent in eluent

Condition system with
820 heating switched
on

Check connections
and make leakproof

Ensure better closure
of eluent supply
vessel, stir

Considerable pressure
drop

Leak in System

Check connections
and make leakproof

Considerable pressure
rise

Contamination of the
filter in the 6.2821.100
Filter unit PEEK

Replace 6.2821.130
connector with filter

Pump is not recognized
in «IC Net»

No RS connection

Switch instrument on.

Check wiring.

Check that the
connection specified in
the 818 IC Pump
window (see section
3.3
Links) matches
the actual wiring