6 appendix, 1 technical specifications, Appendix – Metrohm 818 IC Pump User Manual

Page 42: Technical specifications, N 6 appendix

Advertising
background image

6 Appendix

818 IC Pump / 8.818.1023 Instructions for Use

36

6 Appendix

6.1 Technical

specifications

Provided that nothing to the contrary is mentioned, the

published data are typical values for the 818 IC Pump at

an ambient temperature of 25°C.

Type

Serial dual piston with two valves

Pump capacity

Flow range

0.05...5.00 mL/min

Flow accuracy

< 1.0 %

Flow constancy

< 0.5 % (after unit warmed up)

Reproducibility of

eluent flow

< 0.1 %

Pressure

Pressure range

0...30.0 MPa (0...300 bar)

Resolution for pressure

indicator

0.1 MPa (1 bar)

Pressure measurement

Piezoresistive measurement principle

Response time: 3 ms

Measurement volume: ca. 50

µ

L

Accuracy of pressure

measurement

< 3 %

Residual pulsation

< 1 %

Recorder output for

pressure indication

1 mV/MPa (1 mV/10 bar)

Time constant switchable: HIGH = 20 ms

LOW = 1 s

Safety shutdown

Function

Automatic shutdown when upper and lower

pressure limits violated

Maximum pressure limit

Adjustable between 0.1...30.0 MPa (1...300 bar)

Response time: 1 pump cycle

Minimum pressure limit

Adjustable between 0.1 ... 30.0 MPa (1...300 bar),

inactive at 0 bar

Response time: 3 pump cycles

The cutoff function first becomes active 2 min

after the pump flow has been switched on.

Advertising